Instructions for Use Oxylog 2000 plus SW 1.n 57
Operation
Shutdown
After disconnecting the patient
Switch the ventilator OFF:
1 Press the key
O (A) for approximately
3 seconds. The yellow lamp flashes and
ventilation is terminated by the device.
2 Press the rotary knob (B) to acknowledge the
alarm !!! Confirm device OFF with rotary
knob.
When O
2 is supplied from a cylinder:
3 Close the cylinder valve.
When medical gas is supplied from the pipeline
system:
4 Disconnect the high pressure connection from
the source.
035
WARNING
The cylinder valve must be closed completely
to avoid gas flow leakage by the device.
Oxylog 2000 plus
A
B