Monitoring
66
Instructions for Use Oxylog 2000 plus SW 1.n
Displaying the airway pressure
The actual screen display may differ in appearance
or configuration.
The airway pressure is displayed in a bar graph
indicator on the display.
Refer to the "Operating concept" section on
page 24 for additional information.
Displaying MVe and VTe
MVe and VTe are displayed in the measured MVe /
VTe window.
Refer to the "Operating concept" section on
page 24 for additional information.
To switch between the values:
z Press the upper Values key: the next
value is displayed on the screen.
Displaying O2 values
The O2 concentration can be displayed in the mea-
sured values window.
When the O
2 AirMix – 100% O2 switch has been
changed, the calculated value will be updated after
approximately 30 seconds.
Refer to "Values window" on page 24 for additional
information.
NOTE
This value is a calculated value, based on the
measured air intake and total flow. It is not based
on a measurement by an O
2 sensor!