Assembly
32
Instructions for Use Oxylog 2000 plus SW 1.n
Connecting the power supply
The Oxylog 2000 plus is designed to operate on
power supplies with different voltages.
Internal supply
With rechargeable battery (specified Smart
Battery, refer to the "Technical Data"
on page 103 for additional information).
Additional external power supply
To recharge the battery and to extend the electrical
operation time.
DC voltage from the on-board power supply via
DC/DC converter or
with AC/DC power pack
To have a fully charged battery on hand, refer to the
"Getting started" section on page 38 for additional
information.
Internal supply with rechargeable battery
Replacing the battery
1 Loosen the screw (C) on the battery compart-
ment cover (B) counterclockwise to release the
cover.
2 Remove the battery cover.
3 Remove the battery (A) by pulling the tab.
Checking the charge of the battery
z Press the button on the rechargeable battery.
The charge status is indicated as a percentage
by LEDs.
WARNING
A fully charged battery must always be
installed for safety reasons, even when oper-
ating from an external power supply!
WARNING
Treatment of batteries:
Do not throw into fire,
Do not force open,
Danger of bodily injury.