Configuration
74
Instructions for Use Oxylog 2000 plus SW 1.n
Set startup settings
Range of the settings:
The default settings for the parameters are dis-
played on the screen when the ventilator is
switched ON. The settings can be adjusted.
Advance to the second page:
1 Select the line Page, confirm and turn rotary
knob.
To restore the manufacturer's defaults:
2 Select and confirm line Set factory default.
Select hose type
The type of ventilation hose (reusable or
disposable hose system) can be configured.
The gas output sockets for the flow measuring
hoses must correspond to the hose type system
selected.
Parameter Range
Trigger 0 (OFF) 3 to 15 L/min
PEEP 0 to 20 mbar
I:E 3:1 to 1:4
Ti 0.2 to 10.0 s
Tplat % 0 to 50%
∆Psupp 0 to 35 mbar
Slope SLOW, STANDARD, FAST
NIV ON, OFF
Tapn 0 (OFF), 15 to 60 s
VTapn 50 to 2000 mL
RRapn 12 to 60 bpm
MV-high 2.0 to 41 L/min
MV-low 0.5 to 40 L/min
RR 10 to 100 bpm
Loudness 1/4 to 4/4
Brightness 1/4 to 4/4
screen 20
screen 21screen 24