Operation
50
Instructions for Use Oxylog 2000 plus SW 1.n
VC-SIMV (optional PS)
Volume Controlled - Synchronized Intermittent
Mandatory Ventilation
For patients with inadequate spontaneous breath-
ing, or for patients who are to be weaned gradually.
Fixed mandatory minute volume MV is set with tidal
volume VT and ventilation respiratory rate RR. The
patient can breathe spontaneously between the
mandatory ventilation strokes and thus contribute
to the total minute volume. Spontaneous breathing
can be assisted with PS.
Set the ventilation pattern with the controls below
the display:
Tidal volume VT.
Respiratory Rate RR.
(minimum possible respiratory rate: 2 per min).
Maximum airway pressure Pmax.
–O
2 setting FiO2.
The following are set on the display:
Inspiration time Ti.
Plateau time Tplat %, in % of the inspiration
time.
Positive end expiratory pressure PEEP.
Sensitivity Trigger.
Successful patient triggering is indicated by an
asterisk (*) in the center of the status and alarm
message window.
When setting the ventilation respiratory rate RR,
tidal volume VT or inspiration time Ti, the associ-
ated values for inspiration flow time ratio I:E are
automatically displayed in the information window.
027
Paw
Flow
Ti
Pmax
PEEP
rapid
rise time
slow
rise time
Pressure
support PS
Trigger window
RR
Insp. Flow
t
t
1
screen 9screen 10

Table of Contents