Assembly
30
Instructions for Use Oxylog 2000 plus SW 1.n
When using a bacterial filter or HME
z Connect the bacterial filter or HME to the angled
connector.
Hose connections
1 Connect the ventilation hose (A) to the
breathing valve.
2 Connect the flow measuring hoses (B) to the
nozzles on the flow sensor. Note the different
diameters.
3 Connect the flow measuring (A) hoses to the
Oxylog 2000 plus.
4 Connect the ventilation hose (B) to the gas out-
put on the Oxylog 2000 plus.
NOTE
When using a bacterial filter or HME, measured
flows may deviate from the expiratory flows, as
temperature and humidity of the gas are reduced.
010
WARNING
Bacterial filters increase the exhalation resis-
tance and dead space volume of the ventila-
tion system.
011
A
B
B
CAUTION
Do not use electrically conductive hoses!
Risk of electric shock.
This can endanger the patient.
012
B
A