Problem Solving
84
Instructions for Use Oxylog 2000 plus SW 1.n
Messages in the information window
(Numerical examples)
Message Cause Explanation/Remedy
RR = 12 per min or
VT = 800 mL
I : E = 1 : 1.5
Flow = 15 L/min
Change in Ti, RR or VT in ventila-
tion mode VC-SIMV.
RR = 12 per min or
VT = 800 mL
Ti = 0.7 s
Flow = 35 L/min
Change in I/E, RR or VT in ventila-
tion mode VC-CMV, VC-AC.
I : E = 1 : 1.5
Ti = 2 s
Change in Ti.
PEEP >10 mbar has been set but
not confirmed.
The required setting of
PEEP >10 mbar is only possible
when confirmed via the rotary
knob.
Gas consumption = 10 L/min Standard display in information
window for the current gas
consumption.
(Battery capacity)
Standard display in information
window for the current battery
capacity.
Psupp = 22 mbar Change in Psupp or PEEP. Psupp
is the absolute pressure
resulting from PEEP + Psupp.
Confirm PEEP above 10 mbar ?

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plu and is the answer not in the manual?

Dräger Oxylog 2000 plu Specifications

General IconGeneral
CategoryMedical Equipment
TypeTransport Ventilator
Tidal Volume50 to 2000 mL
PEEP0 to 20 cmH2O
DisplayMonochrome LCD
Inspiratory Time0.2 to 5 s
FiO221 to 100%
Power Supply100 to 240 V AC, 50/60 Hz
Ventilation ModesIPPV, SIMV, CPAP

Related product manuals