Problem Solving
84
Instructions for Use Oxylog 2000 plus SW 1.n
Messages in the information window
(Numerical examples)
Message Cause Explanation/Remedy
RR = 12 per min or
VT = 800 mL
I : E = 1 : 1.5
Flow = 15 L/min
Change in Ti, RR or VT in ventila-
tion mode VC-SIMV.
RR = 12 per min or
VT = 800 mL
Ti = 0.7 s
Flow = 35 L/min
Change in I/E, RR or VT in ventila-
tion mode VC-CMV, VC-AC.
I : E = 1 : 1.5
Ti = 2 s
Change in Ti.
PEEP >10 mbar has been set but
not confirmed.
The required setting of
PEEP >10 mbar is only possible
when confirmed via the rotary
knob.
Gas consumption = 10 L/min Standard display in information
window for the current gas
consumption.
(Battery capacity)
Standard display in information
window for the current battery
capacity.
Psupp = 22 mbar Change in ∆Psupp or PEEP. Psupp
is the absolute pressure
resulting from PEEP + ∆Psupp.
Confirm PEEP above 10 mbar ?