Instructions for Use Oxylog 2000 plus SW 1.n 85
Problem Solving
Error messages during the device check
Message Cause Explanation/Remedy
No communication control- /
charge-board
Device defective Contact your local DrägerService
for additional support.
System leakage Leak in ventilation hose system
and/or test lung.
Check hoses, breathing valve,
flow sensor and test lung for leaks
and replace if necessary.
No test lung Test lung not connected or mayor
leakage
Connect test lung.
Check hoses, breathing valve,
flow sensor and test lung for leaks
and replace if necessary.
Breathing valve inop PEEP >10 mbar has been set but
not confirmed.
The required setting of
PEEP >10 mbar is only possible
when confirmed via the rotary
knob.
Pressure measurement inop The ventilation hose system has
not been connected correctly.
Connect ventilation system cor-
rectly.
Pressure measurement is not pos-
sible.
Contact your local DrägerService
for additional support.
PEEP-valve inop Internal leak in system Check hoses, breathing valve,
flow sensor and test lung for leaks
and replace if necessary.
Device defective Contact your local DrägerService
for additional support.
Patient flow measurement
inop
Flow measurement implausible Replace flow sensor.
Contact your local DrägerService
for additional support.