Cleaning, Disinfection and Sterilization
90
Instructions for Use Oxylog 2000 plus SW 1.n
Remove the disposable hose system
1 Disconnect flow measuring hoses (A and B).
2 Disconnect the ventilation hose (C).
3 Correctly dispose of the complete disposable
hose system. Refer to the chapter "Disposal"
on page 99.
WARNING
The rubber disc (A) in the housing must not be
removed, damaged or bent, otherwise the
valve will not work properly and endangers
the patient.
Risk of CO2 rebreathing.
047
CAUTION
The disposable hose system must not be
sterilized: it cannot withstand high temperatures
and may be damaged!
A
B
C