Instructions for Use Oxylog 2000 plus SW 1.n 89
Cleaning, Disinfection and Sterilization
4 Disconnect the flow sensor (A) from the breath-
ing valve.
5 Carefully detach the flow measuring hoses (B)
from the flow sensor.
Pull in the axial direction of the hose nozzles.
6 Detach the angled connector (C) from the flow
sensor.
7 Detach the ventilation hose (D) from the brea-
thing valve.
Breathing valve, disassembly
8 Turn the cover (A) about 90
o
counterclockwise to unlock the cover.
9 Remove the silicone diaphragm (B).
011
WARNING
Do not twist or use force when disconnecting
the flow measuring hoses from the flow sen-
sor nozzles. This can damage the flow sensor.
WARNING
Do not allow any objects to enter the flow sen-
sor to prevent risk of malfunctions.
Do not purge with compressed air. The wind
vane inside may be damaged and cause mea-
suring errors!
B
A
C
D
B
011
WARNING
Do not disassemble breathing valve any fur-
ther.
Do not allow any objects to enter the housing
of the breathing valve to prevent risk of mal-
functions!
Do not damage the silicone diaphragm and
other parts.
011
A
B
A