Technical Data
110
Instructions for Use Oxylog 2000 plus SW 1.n
AC/DC power pack
Temperature range –20
o
C to 50
o
C
Protection class to
EN 60601
Class II, the earthing is used for EMC purposes.
Input 100 to 240 V~/ 50 to 60 Hz / 0.6 - 0.3 A
Output 19 V ±0.5 V / 2.1 A
DC/DC converter
Temperature range –20
o
C to 50
o
C
Input 12 / 24 / 28 V DC / 9 A
Output 19 V ±0.5 V / 2.1 A
Gas supply
From a pipeline system or from a medical gas
cylinder.
O
2 service pressure 270 kPa to 600 kPa maximum 100 L/min
Supply gas Medical oxygen.
O
2 inlet connection either:
NIST
1)
to EN 739, or
DISS
2)
to CGA V5-1989, or
N-F
3)
S90-116
WARNING
Only use medical grade oxygen that is dry and
free from dust and oil.
Contaminated gas can cause device malfunction.
Gas cylinders and
pressure reducers
WARNING
Only use compressed gas cylinders and pres-
sure reducers, which comply with all applicable
regulations and have been approved.
Pressure reducer Must have a vent valve on the output side to limit the
delivery pressure to approximately 1000 kPa in the
event of a fault.
Gas consumption for
internal control
Average 0.5 L/min
Accuracy of gas
consumption indication
15% or ±1 L/min, whichever is greater.
1) NIST = Non Interchangeable Screw Thread Connection
2) DISS = Diameter Index Safety Systems
3) N-F = French standard