Instructions for Use Oxylog 2000 plus SW 1.n 41
Getting Started
Perform device check
The device check consists of the following steps.
– Duration is approximately 3 minutes.
Connect the test lung
1 Connect the angled adapter (A) to the
breathing valve.
2 Connect the catheter connector (B), diameter
7 mm, to the angled adapter.
The catheter connector simulates the
resistance of the airways.
3 Connect the test lung (C).
Switch ON
1 To switch ON briefly press the O key (B).
The device performs a self-test and the
operator is prompted, on the display, to activate the
configuration menu or device check:
Press rotary knob for device check and
configuration
2 Press the rotary knob (A) to confirm, before the
bar is full.
3 Select Device check in the main menu and
confirm.
021
CAUTION
BTPS values of a test lung are not the same as the
BTPS values of a patient. The Oxylog 2000 plus
measures and adapts according to BTPS values of
a patient. Therefore, when a test lung is con-
nected, the MVe and VTe indicated on display may
differ from the MVe and VTe that is set by the
operator.
A
C
B
022screen 6
NOTE:
The device check can be discontinued at any time
by pressing the Alarm Reset key.
Oxylog 2000 plus
A
B