Instructions for Use Oxylog 2000 plus SW 1.n 39
Getting Started
Determining the approximate pneumatic operating time for the
Oxylog 2000 plus
Example for supply of medical gas:
– Cylinder pressure measured on the pressure
gauge of the pressure reducer: 2000 kPa
– Liquid capacity of the O
2 cylinder: 2.1 L
Supply of medical gas:
2.1 L x 2000 kPa = approximately 420 L
Example for pneumatic operation time:
– VC-CMV mode, respiratory rate
10 breaths /min, VT = 1 L, O
2 = 100%
– Minute volume = 10 breaths /min x 1 L =
10 L/min
Operation time =
*
Calculated with average gas consumption of
ventilator: 0.5 L/min
Operation time = = approx. 40 minutes
The pneumatic operation time increases when
Oxylog 2000 plus operates with O2 AirMix, as
ambient air is drawn into the device.
The amount of gas from the high-pressure supply,
which is currently being consumed, is indicated by
the Oxylog 2000 plus in the lower left section of the
information window in L/min. This display is over-
written when a higher priority message is activated.
Example:
O
2 consumption = 2.1 L/min
Medical gas supply [L]
(MV +0.5*) [L/min]
420
10.5