Instructions for Use Oxylog 2000 plus SW 1.n 93
Cleaning, Disinfection and Sterilization
– Sterilization for longer than 10 minutes is
permissible, but will decrease the service life of
the hose system.
After care
– Reassemble, refer to the "Assembly" section for
information.
– Connect to the power supply and gas supply,
refer to the "Assembly" section for information.
– Check readiness for operation, refer to the
"Assembly" section for information.
Note service life of the hose system
The parts of the breathing valve, the flow sensor,
the angled connector, the flow measuring hoses
and the ventilation hose are resistant to the recom-
mended disinfectants and to the temperatures
occurring during sterilization.
However, every disinfection and sterilization cycle
also means wear on the parts concerned.
For this reason, the parts must be examined for
cracks and permanent deformation after the
cleaning procedure.
NOTE
Damaged or deformed parts must be replaced.