36
After sterilisation, check the hand grip for mechanical
integrity.
Do not use damaged hand grips.
9. SAFETY INSPECTIONS
EVERY YEAR:
Check the power supply cable for damage and
replace if necessary.
Check for distortions and cracks in the plastic parts.
Check the load-bearing system for distortion or damage.
Check for loose parts.
9.1 SETTING THE SPRING FORCE
CAUTION
The ideal spring force is set in the factory.
Polaris 50 Ceiling / Polaris 50 Wall
CAUTION
The luminaire head must be installed before the
spring force is adjusted.
Move the spring-balanced arm to the highest possible
position.
Insert a 6 mm Allen wrench through the opening on the
middle joint and adjust the spring force by turning the
spring.
Turn the screw clockwise (+) to increase the spring force
(if the spring-balanced arm lowers).
Turn the screws anticlockwise (-) to reduce the spring
force (if the spring-balanced arm rises).
CAUTION
Damage to the hand grip
Do not sterilise with hot air
Wrap the hand grip in a sterile bag before sterilisation.
The hand grip is designed exclusively for damp
sterilisation with pre-vacuum and saturated steam
fractioned 3 times with the following parameters:
Temperature
134°C
Overpressure
2.0 bars
Contact time
6 min
Vacuum drying
20 min
DANGER
Risk of death from electric shock.
Unplug from the grid
The power supply cable must be checked at least
once a year for damage.
CAUTION
Maintenance and repairs should only be performed
by qualified electricians.
The corresponding user profile is in Chapter 2, Safety
instructions.