10 | Channel setup menu 4
D-ISC 100 x xx2
113
With activated software module "MC0: Media conditions", the
media pressure is distributed from D−ISC100 to all connec-
ted sensors.
Standard temperature [°C]
Temperature used for standardisation
If the standardised temperature value differs from 0°C, this
can be set here.
Default value = 0°C
Standard density [kg/m³]
Standard density of the measuring medium.
For example, this is required to calculate the mass flow.
10.3.5.4 [Maintenance setup]
Reference check set value:
Reference check set value
Press the
key to change…
Value in % to which the sensor's internal reference point
measurement is scaled in order to output the value in the set
signal range.
Adjustment range 5…95%
Control cycle step duration:
Control cycle step duration
Press the
key to change…
Setting the duration of each individual step of the check cycle
in seconds.
Check cycle interval:
Control cycle interval
Press the
key to change…
Setting the interval of the automatic check cycle in hours.
0 = automatic check cycle is disabled.
Fault indication rejection:
Fault indication rejection
Press the
key to change…
If a failure is detected by the system, this results in a pro-
grammed response. The system also saves the appropriate
(error) message. The message is saved in the memory of the
message logbook (in the sensor) and can provide the main-
tenance technician with information for/when carrying out
device maintenance, if applicable.
Error message suppression designates the time period (in
seconds) for the sensor's initial pause. The system response
is then only initiated if the fault still persists after the specified
time has expired. This means transient failures can be ig-
nored without triggering a system response.
Time to next control cycle
[h:mm:ss]:
Time to next control cycle [h:mm:ss]
(read only)
Indication of remaining time to next control cycle.
Display -:--:--= automatic check cycle is disabled.
Reset control cycle timer:
Reset control cycle timer
Carry out (!) with the
key.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the DURAG D-ISC 100 2 Series and is the answer not in the manual?

DURAG D-ISC 100 2 Series Specifications

General IconGeneral
BrandDURAG
ModelD-ISC 100 2 Series
CategoryControl Unit
LanguageEnglish

Related product manuals