15 | Examples of settings
200
D-ISC 100 x xx2
15.16 Example: Configuring the Media condition
software module
In our 15.14 Example: Configuring the mixed channel software
module [}191], we wanted to determine the volume of dust flow-
ing out of a stack.
The [
Media condition] software module makes it possible to
record and display the media conditions for the measured me-
dium (temperature and absolute pressure) for the entire meas-
urement system, and to provide these for all connected DURAG
sensors in order to standardise the measured values.
Channel #1 of module MC0 is normally used for the media tem-
perature
(MC0.1).
Channel #2 of module MC0 is normally used for the media pres-
sure
(MC0.2).
Requirement:
The [
Media condition] software module must be enabled.
see 10.4.1.3 Enable/disable modules [}116].