EasyManua.ls Logo

Ecotech Serinus 40 - Page 108

Ecotech Serinus 40
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Serinus 40 User Manual 2.2
Page 108
2. Disconnect the electrical plug and the two tube connections from the assembly. Remove the
four screws on the side (remove the coil and ozone generator cell assembly after ensuring the
pneumatic and electrical connections are disconnected).
Figure 63 Ozone generator cell assembly
Figure 64 Removing ozone generator cell assembly
3. Remove the ozone producing cell assembly by first unscrewing the spring conductor connection
and then removing the screws that secure the Teflon blocks at each end of the assembly.
CAUTION
Take care not to knock glass tube.
4. Disassemble the glass and shaft from the Teflon blocks gently separating them by hand.
5. Remove the four O-rings. During services it is recommended that the stainless steel shaft
conductor is cleaned and polished with alcohol solution. If heavily oxidized, polish with steel wool
or fine grade wet and dry sandpaper. Ensure any finger marks are wiped clean.