Service
Page 109
Figure 65 Disassembled ozone generator cell assembly
6. Fit the new Viton O-rings (refer to Figure 66) O010019 and O010020.
Figure 66 Replacing Viton O-rings
7. Reassemble the ozone generator assembly. A complete assembly leak check should be
performed before re-installing the unit in the analyser.