VE-1010 Veterinary PC ECG User Manual Safety Guidance
- 4 -
CAUTION
e) Test the protection earth resistance according to IEC/EN 60601-1: Limit: 0.1 ohm.
f) Test the enclosure leakage current according to IEC/EN 60601-1: Limit: NC 100
μA, SFC 500 μA.
g) Test the patient leakage current according to IEC/EN 60601-1: Limit: NC a.c. 10
μA, d.c. 10 μA; SFC a.c. 50 μA, d.c. 50 μA.
h) Test the patient auxiliary current according to IEC/EN 60601-1: Limit: NC a.c. 10
μA, d.c. 10 μA; SFC a.c. 50 μA, d.c. 50 μA.
i) Test the patient leakage current under single fault condition with mains voltage on
the applied part according to IEC/EN 60601-1: Limit: 50 μA (CF).
The data should be recorded in an equipment log. If the equipment is not functioning
properly or fails any of the above tests, the equipment has to be repaired.
8. The device and accessories are to be disposed of according to local regulations after
their useful lives. Alternatively, they can be returned to the dealer or the
manufacturer for recycling or proper disposal.
9. Precautionary maintenance of this system, including periodically cleaning and
checking the appearance, can be finished by users because this maintenance does
not touch the interior.
10. Before cleaning or maintaining the system, turn off the system power and remove
the power cable.
11. Prevent the detergent from seeping into the equipment while cleaning.
12. Avoid pouring liquid on the equipment while cleaning, and do not immerse any parts
of the equipment into any liquid.
13. Do not clean the unit and accessories with abrasive fabric and avoid scratching the
electrodes. Removing all dust from the exterior surface of the equipment with a soft
brush or cloth, or with a soft cloth which is slightly dampened with a mild detergent
solution or cool disinfector. Especially the tie-in and the panel edge should be
noticed.
14. Any remainder of detergent should be removed from the unit and the patient cable
after cleaning.
15. Do not use chloric disinfectant such as chloride, sodium hypochlorite etc.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN INSTRUMENTS VE-1010 and is the answer not in the manual?

EDAN INSTRUMENTS VE-1010 Specifications

General IconGeneral
BrandEDAN INSTRUMENTS
ModelVE-1010
CategoryMedical Equipment
LanguageEnglish

Related product manuals