VE-1010 Veterinary PC ECG User Manual Technical Specifications
- 64 -
Appendix 1 Technical Specifications
A1.1 Safety Specifications
Comply with:
IEC 60601-1: 1988+A1+A2, EN 60601-1:1990+A1+A2,
IEC/EN60601-1-2: 2001+A1, IEC/EN60601-2-25,
ANSI/AAMI EC11, IEC/EN60601-2-51
Anti-electric-shock type: Class ІI
Anti-electric-shock degree: Type CF with defibrillation-proof
Degree of protection against
harmful ingress of water:
Ordinary equipment (Sealed equipment without liquid proof)
Disinfection/sterilization
method:
Refer to the user manual for details
Degree of safety of application
in the presence of flammable
gas:
Equipment not suitable for use in the presence of flammable
gas
Working mode: Continuous operation
EMC: Group І, Class A
NC <10μA (AC) / <10μA (DC)
Patient
Leakage
Current:
SFC <50μA (AC) / <50μA (DC)
NC <10μA (AC) / <10μA (DC)
Patient
Auxiliary
Current:
SFC <50μA (AC) / <50μA (DC)
A1.2 Environment Specifications
Transport & Storage Working
Temperature:
DP12 ECG sampling box:
-40ºC (-40ºF) ~ +55ºC (+131ºF)
+5ºC (+41ºF) ~ +40ºC (+104ºF)
Relative Humidity:
25%~93%
Non-Condensing
25%~80%
Non-Condensing
Atmospheric Pressure: 700hPa ~1060hPa 860hPa ~1060hPa