VE-1010 Veterinary PC ECG User Manual Introduction
- 10 -
WARNING
1. Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing
equipment and IEC/EN 60601-1 for medical equipment). Furthermore all
configurations shall comply with the valid version of the standard IEC/EN 60601-1-1.
Therefore anybody, who connects additional equipment to the signal input or output
connector to configure a medical system, must make sure that it complies with the
requirements of the valid version of the system standard IEC/EN 60601-1-1. If in
doubt, consult our technical service department or your local distributor.
2. The summation of leakage current should never exceed the leakage current limit
while several other units are used at the same time.
2.2 Features
Powerful functions, friendly interfaces and easy operation
6/7-channel ECG waves are displayed and printed simultaneously
ECG waves can be frozen and reviewed
Measurement point adjustment and re-analysis, manual measurement with an electronic ruler
of high precision
Perfect data management and processing functions
Reports can be printed in PDF, Word, JPG or BMP format
Supporting multi-language
Supporting auto measurement
Automatic baseline adjustment for optimal printing
High performance filters guarantee stable ECG waveforms
Real-time analysis, real-time displaying and printing 6 or 7 leads simultaneous ECG
waveforms
Supporting editing diagnosis template
Two analysis functions including Normal ECG and HRV analysis