2. Isolate the power and disconnect the electrical supply plug from the electrical
supply connect
or, then isolate the water and the gas purge supply.
3. Disconnect the purge gas supply, taking care as any trapped gas under pressure is
released. Disconnect the cooling water supply followed by the cooling water
return.
4. Disconnect the inlet and outlet from the vacuum and exhaust systems, remove
bulky accessories such as lters, silencers and knock out pots (if installed) and then
t the blanking caps.
5. If necessary, disconnect any other accessories from the dry pumping system.
6. For dry pumping systems with castors, adjust the levelling feet so that the dry
pumping system rests on the castors.
7. Move the dry pumping system to the locaon where maintenance will be carried
out.
Aer maintenance is complete, re-install the dry pumping system as described in
Installaon on page 27.
7.4 Draining the cooling water
WARNING:
Use of suitable pr
otecve gloves and eye protecon is recommended when carrying
out this procedure. Personal protecve equipments should be checked and used as
specied by its supplier.
1. Relocate the dry pumping system for maintenance as stated in Relocate the dry
pumping system for maintenance on page 57.
2. Connect a regulated clean dry air supply (5 barg or 73 psig) to the cooling water
supply connecon (Refer to Rear view on page 12). Do not turn on the air supply
yet.
3. Connect a drain hose to the cooling water return connecon (R
efer to Rear view on
page 12). Posion the open end of the drain hose in a suitable collecon container.
4. Turn on the clean dry air supply.
5. Monitor the drain hose outlet unl no further cooling water is purged.
6. The cooling water drain procedure is now complete. Disconnect the air supply,
electrical supply and drain hoses. Dispose of the drained cooling water
appropriately.
7.5 General maintenance
CAUTION:
Do not use cleaning mat
erials based on strong alkalis, aggressive or chlorinated
solvents. Do not use cleaning materials containing abrasives.
The following maintenance can be carried out on the dry pumping system between the
overhauls. Contact us for the details and training:
▪ Check the oil level. Refer to Checking the oil levels and relling on page 59.
07/2021 - ©Edwards Limited