EasyManua.ls Logo

Edwards nEXT730 - Page 26

Edwards nEXT730
137 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
B8J200880_E - Technical data
Page 25
nEXT
730D
730D
930D
930D
730H
730H
730Q
930Q
Power consumption at ultimate pressure
W
40
Type of protection
IP
Type 1 (UL
50E)
54
1
Cooling standard
Convection*
Water*
Cooling optional
Air or water*
-
Cooling water connection
plug-in connection for 6x1 hose / alternative G 1/8"
Cooling water consumption
l/h
60
Permissible cooling water pressure (bar(g):
bar (gauge) is overpressure, i.e. atmospheric
pressure = 0 bar(g))
bar(g)
6
Permissible cooling water temperature
°C
15 to 35
Weight
kg
14.6
19.6
15.4
21.7
14.6
19.6
15.4
Recommended forevacuum pumps
XDS35i or E2M28
Noise level
with convection cooling
with radial air cooler
dB(A)
< 40
< 55
< 40
-
Maximum bake-out temperature of the CF
version, -water cooled
°C
100
-
Purge gas flow
mbar · I · s
-1
sccm
0.4
24
Purge gas connection
G 1/8"
Venting connection
G 1/8"

Table of Contents