B8J200880_E - Technical data
Page 46
Object name
Command
Parameter range
Factory
setting
Data type
Units
Comments
Normal speed
setting
!S856
50..100
80
decimal
%
Normal speed as a percentage of full
speed
?S856
-
-
-
-
-
Standby speed
setting
!S857
55..100
70
decimal
%
Standby speed as a percentage of full
speed
?S857
-
-
-
-
-
Temperature
readings
?V859
0..100
-
decimal
°C
Measured motor temperature
0..100
decimal
°C
Measured motor controller tempera-
ture
Link parameter
readings
?V860
0..500
-
decimal
0.1 Volts
Measured link voltage
0..300
decimal
0.1 Amps
Measured link current
0..15000
decimal
0.1 Watts
Measured link power
Pump run hours
?V862
0..65535
-
decimal
hours
Hours run by pump
Temperature
readings 2
?V865
-200..200
-
decimal
°C
Measured motor temperature
-200..200
decimal
°C
Measured motor controller tempera-
ture
-200..200
decimal
°C
Measured bearing temperature
Factory settings
!S867
1
-
-
-
Reset all configuration options and
parameters to the factory settings
PIC software version
?S868
10
-
string
chars
Boot loader software version number
(D396696XXX)
Speed control
!C869
0
-
decimal
-
Set target speed to full speed
1
-
-
-
Set target speed to standby speed

Table of Contents