Operating faults
© Elma Schmidbauer GmbH BA_Elmasolvex_VA_EN_Vers.05.2018
95
267
Abortion
Critical fault!
Working level filling
level sensor defective.
Malfunction of the Working
level fill level sensor during
the safety test.
Check the following possibilities:
Cleaning chamber empty (correct)?
Working level fill level sensor
(Fig.9.1.4.1.A) soiled? Clean with
suitable cloth if necessary.
Work level fill level sensor
(Fig.9.1.4.1.A) damaged (scratches,
cracks)?
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
268
Abortion
Critical fault!
Vacuum is not
achieved.
Vacuum cannot be achieved
during the safety test.
Check the following possibilities:
Is the cover of the cleaning chamber
closed correctly?
Cover seal of the cleaning chamber
(chapter 10.1.1.B) damaged / slipped?
Emergency ventilation in the cover of
the cleaning chamber (chapter
11.3.2.A) leaking?
Condensate separator, activated
carbon filter (if connected) clogged?
Exhaust air tube (chapter 4.3) kinked?
Can the valves of the quick couplings
(Fig. 4.4.1.2.J) be pressed in / moved
correctly?
Can the valves of the quick couplings
on the media tanks (Fig. 4.4.1.1.F) be
pressed in / moved correctly?
Are the quick couplings on the media
tanks (Fig. 3.13.1. B/C) correctly
engaged?
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
269
Abortion
Critical fault!
Slow venting not
possible.
Vacuum in the cleaning
chamber is not released
during the safety test.
Check the following possibilities:
Air intake and exhaust openings in the
cleaning chamber (Fig. 3.12.1.A)
clogged?
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
270
Abortion
Critical fault!
Fast venting not
possible.
Vacuum in the cleaning
chamber is not released
during the safety test.
Check the following possibilities:
Air intake and exhaust openings in the
cleaning chamber (Fig. 3.12.1.A)
clogged?
Switch off cleaning machine and
restart: Contact Service if the fault
persists.