Operating faults
96
BA_Elmasolvex_VA_EN_Vers.05.2018 © Elma Schmidbauer GmbH
271
Abortion
Critical fault!
Basket rotation drive
defective.
Basket rotation drive
communication error and
speed fault during the safety
test.
Check the following possibilities:
Check correct seating of the cleaning
basket
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
272
Abortion
Critical fault!
Ultrasound generator
defective.
Ultrasound generator
communication error during
the safety test.
Check the following possibilities:
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
273
Abortion
Critical fault!
Fan defective.
A fan is blocked or defective
during the safety test.
Check the following possibilities:
Check fans for foreign bodies and
remove if necessary
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
274
Abortion
Critical fault!
Both fans defective.
Both fans are blocked or
defective during the safety
test.
Check the following possibilities:
Check fans for foreign bodies and
remove if necessary
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
278
Abortion
Critical fault!
Memory chip defective.
Software / electronics error. Check the following possibilities:
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
291
Abortion
Critical fault!
Basket drive defective
or blocked.
Basket rotation drive
defective or blocked.
Check the following possibilities:
Check motor shaft of the basket
rotation drive for ease of movement.
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
292
Warning
Critical fault!
Ultrasound generator
defective.
No communication with
ultrasound generator.
Check the following possibilities:
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
297
If liquid is detected at the
Check the following possibilities: