Description of parameters (FF configuration program) Cerabar S/Deltabar S
52 Endress+Hauser
6.3.3 Service Transducer Block
SENSOR_
MEASUREMENT_TYPE/
SENSOR MEAS. TYPE
Display
Index: 90
Data type: Unsigned16
Access: Read only
Displays the sensor type.
Deltabar S = differential
Cerabar S with gauge pressure sensors = relative
Cerabar S with absolute pressure sensors = absolute
Service Transducer Block
Parameter Description
CONFIGURATION
COUNTER/
CONFIG. RECORDER
Display
Index: 12
Data type: Unsigned16
Access: Read only
Displays the configuration counter.
This counter is increased by one with each change to a manufacturer-specific parameter. The
counter counts to 65535 and then starts again at zero. Changes to parameters that configure
the local display, such as the DISPLAY_ LANGUAGE/DISPLAY LANGUAGE parameter, do
not increase the counter.
MEASURED_
TEMPERATURE_3/
TEMP. SENSOR
Display
Index: 13
Data type: Float
Access: Read only
Displays the measured temperature of the main electronics.
TEMPERATURE_3_LOW
_LIMIT/
ALLOWED MIN. TEMP
Display
Index: 14
Data type: Float
Access: Read only
Displays the lower temperature limit of the main electronics.
TEMPERATURE_3_
HIGH_LIMIT/
ALLOWED MAX. TEMP
Display
Index: 15
Data type: Float
Access: Read only
Displays the upper temperature limit of the main electronics.
PROCESS_CONNEC-
TION_PRESSURE_HIGH
_ LIMIT/
Pmax PROC. CONN.
Entry
Index: 16
Data type: Float
Access: Auto, OOS
For entering and displaying the maximum permitted pressure of the process connection.
Factory setting:
In accordance with nameplate data (see also Operating Instructions BA301P (Deltabar S)
or BA302P (Cerabar S), Section 2.1.1 nameplate)
Pressure Transducer Block
Parameter Description