Description of parameters (FF configuration program) Cerabar S/Deltabar S
62 Endress+Hauser
DIAGNOSTIC_CODE/
ACTUAL ERROR
Display
Index: 15
Data type: Unsigned16
Access: Read only
Displays the current messages present. If two or more messages are present, the message
with the highest priority is the one shown on the display. → See also these Operating
Instructions, Section 11.1. "Messages" and Section 11.3 "Confirming messages".
LAST_DIAGNOSTIC_
CODE/
LAST DIAG. CODE
Display
Index: 16
Data type: Unsigned16
Access: Read only
Displays the last error that occurred and was eliminated.
!
Note!
• The message displayed here can be cleared by means of the RESET_ALARM_HISTORY/
RESET ALL ALARMS parameter.
ACKNOWLEDGE_
ALARM_MODE/
ACK. ALARM MODE
Selection
Index: 17
Data type: Unsigned8
Access: Auto, OOS
Switch on acknowledge alarm mode.
→ See also ACKNOWLEDGE_ALARM/ACK. ALARM.
Options:
•On
•Off
Factory setting:
Off
ACKNOWLEDGE_
ALARM/
ACK. ALARM
Selection
Index: 18
Data type: Unsigned8
Access: Auto, OOS
Acknowledge alarm.
Prerequisite:
• ACKNOWLEDGE_ALARM_MODE/ACK. ALARM MODE = On
Options:
• Abort
•Confirm
The cause of the alarm must be eliminated, the message must be acknowledged via the
ACKNOWLEDGE_ALARM/ACK. ALARM parameter and, where applicable, the
ALARM_HOLD_TIME/ALARM DISPL. TIME (→ Page 63) has to have elapsed before the
device starts measuring again following an alarm.
→ See also these Operating Instructions, Section 11.3 "Confirming messages".
Factory setting:
Abort
RESET_ALARM_
HISTORY/
RESET ALL ALARMS
Selection
Index: 19
Data type: Unsigned8
Access: Auto, OOS
Use this parameter to reset the message of the LAST_DIAGNOSTIC_CODE/LAST DIAG.
CODE parameter.
Options:
• Abort
•Confirm
Factory setting:
Abort
REACTION_ON_
ALARM_ NR/
ERROR NO.
Entry
Index: 20
Data type: Unsigned16
Access: Auto, OOS
For "Error"-type messages, you can decide whether the device should behave as in the event
of an alarm (A) or as in the event of a warning (W). Enter the corresponding message number
for this parameter. → See also CHOOSE_ERROR_OR_WARNING/SELECT ALARM TYPE.
→ See also these Operating Instructions, Section 11.1 "Messages" and Section 11.2
"Response of outputs to errors".
Diagnosis Transducer Block
Parameter Description