Description of parameters (FF configuration program) Cerabar S/Deltabar S
62 Endress+Hauser
DIAGNOSTIC_CODE/
ACTUAL ERROR
Display
Index: 15
Data type: Unsigned16
Access: Read only
Displays the current messages present. If two or more messages are present, the message
with the highest priority is the one shown on the display. See also these Operating
Instructions, Section 11.1. "Messages" and Section 11.3 "Confirming messages".
LAST_DIAGNOSTIC_
CODE/
LAST DIAG. CODE
Display
Index: 16
Data type: Unsigned16
Access: Read only
Displays the last error that occurred and was eliminated.
!
Note!
The message displayed here can be cleared by means of the RESET_ALARM_HISTORY/
RESET ALL ALARMS parameter.
ACKNOWLEDGE_
ALARM_MODE/
ACK. ALARM MODE
Selection
Index: 17
Data type: Unsigned8
Access: Auto, OOS
Switch on acknowledge alarm mode.
See also ACKNOWLEDGE_ALARM/ACK. ALARM.
Options:
•On
•Off
Factory setting:
Off
ACKNOWLEDGE_
ALARM/
ACK. ALARM
Selection
Index: 18
Data type: Unsigned8
Access: Auto, OOS
Acknowledge alarm.
Prerequisite:
ACKNOWLEDGE_ALARM_MODE/ACK. ALARM MODE = On
Options:
Abort
•Confirm
The cause of the alarm must be eliminated, the message must be acknowledged via the
ACKNOWLEDGE_ALARM/ACK. ALARM parameter and, where applicable, the
ALARM_HOLD_TIME/ALARM DISPL. TIME (Page 63) has to have elapsed before the
device starts measuring again following an alarm.
See also these Operating Instructions, Section 11.3 "Confirming messages".
Factory setting:
Abort
RESET_ALARM_
HISTORY/
RESET ALL ALARMS
Selection
Index: 19
Data type: Unsigned8
Access: Auto, OOS
Use this parameter to reset the message of the LAST_DIAGNOSTIC_CODE/LAST DIAG.
CODE parameter.
Options:
Abort
•Confirm
Factory setting:
Abort
REACTION_ON_
ALARM_ NR/
ERROR NO.
Entry
Index: 20
Data type: Unsigned16
Access: Auto, OOS
For "Error"-type messages, you can decide whether the device should behave as in the event
of an alarm (A) or as in the event of a warning (W). Enter the corresponding message number
for this parameter. See also CHOOSE_ERROR_OR_WARNING/SELECT ALARM TYPE.
See also these Operating Instructions, Section 11.1 "Messages" and Section 11.2
"Response of outputs to errors".
Diagnosis Transducer Block
Parameter Description