Description of parameters (FF configuration program) Cerabar S/Deltabar S
64 Endress+Hauser
TEMPERATURE_UNIT/
TEMP. ENG. UNIT
Display
Index: 27
Data type: Unsigend16
Access: Read only
Displays the temperature unit selected.
The temperature unit is selected by means of the SECONDARY_VALUE_UNIT parameter
(→ Page 35) in the Pressure Transducer Block.
TEMPERATURE_1_
USER_LOW_LIMIT/
Tmin ALARM WINDOW
Entry
Index: 28
Data type: Float
Access: Auto, OOS
Customer-specific process monitoring – enter lower temperature limit.
You can use the REACTION_ON_ALARM_NR/SELECT ALARM TYPE parameter to enter
how the device responds if the operating pressure undershoots the specified value.
→ See also these Operating Instructions, Section 11.1 "Messages", Table, Code E732 and
Section 11.2. "Response of outputs to errors".
Factory setting:
Lower sensor temperature operating limit – 10 K
(→ for the lower temperature operating limit, see TEMPERATURE_1_SENSOR_LIMIT_
LOW/Tmin SENSOR, Page 53)
TEMPERATURE_1_
USER_HIGH_LIMIT/
Tmax ALARM WINDOW
Entry
Index: 29
Data type: Float
Access: Auto, OOS
Customer-specific process monitoring – enter upper temperature limit.
You can use the REACTION_ON_ALARM_NR/SELECT ALARM TYPE parameter to enter
how the device responds if the operating pressure undershoots the specified value.
→ See also these Operating Instructions, Section 11.1 "Messages", Table, Code E733 and
Section 11.2. "Response of outputs to errors".
Factory setting:
Upper sensor temperature operating limit +10 K
(→ for the upper temperature operating limit, see TEMPERATURE_1_SENSOR_LIMIT_
HIGH/Tmax SENSOR, Page 53)
RESET_INPUT_VALUE/
ENTER RESET CODE
Entry
Index: 30
Data type: Unsigned16
Access: Auto, OOS
Reset parameters completely or partially to factory values or delivery status.
→ See also Operating Instructions BA301P (Deltabar S) or BA302P (Cerabar S), Section 5.7
"Factory setting" (reset).
Factory setting:
0
OPERATING_HOURS
Display
Index: 31
Data type: Unsigned32
Access: Read only
Displays the hours of operation. This parameter cannot be reset.
ALARM_TABLE_ARRAY
/ALARM TABLE ENTRY
Display
Index: 32
Data type: Visible String
Access: Read only
This parameter displays pending alarms and warnings in hexadecimal numerical format
(max. 16). Every message is assigned to a bit in the sequence of the error code in accordance
with Section 11.1 "Messages".
Diagnosis Transducer Block
Parameter Description