Gammapilot M FMG60 with FOUNDATION Fieldbus Commissioning
Endress+Hauser 75
"Value full" (*14)
"Value empty" (*17)
These functions are needed for level measurements only. They are used to specify the level at which
the full or empty calibration are performed.
Range of values
"Calibration start stop" (*15)
This function is used to start the automatic entering of the selected calibration point.
Options:
stop/edit This option must be selected if
the calibration point is not to be entered (e.g. because it has already been entered). The pulse
rate of the calibration point is then displayed in the following function, "full calibr." (*16) or
"empty calibr." (*18). If required, this value can be changed.
the calibration point is to be entered manually. This can be done in the following function,
"full calibr." (*16) or "empty calibr." (*18).
start
This option is used to start the automatic entering of the calibration point. The Gammapilot M
then changes to the "avg. pulse rate" (*11) function.
value full *14
100%
ENDRESS + HAUSER
E
+
value empty *17
0%
ENDRESS + HAUSER
E
+
optimum value minimum value maximum value
Value full (*14) 100% 60% 100%
Value empty (*17) 0% 0% 40%
calibration start stop
*15
stop/edit
start
ENDRESS + HAUSER
E
+