Mounting Liquicap M FMI51 HART
14 Endress+Hauser
4.2 Mounting requirements
4.2.1 Mounting the sensor
The Liquicap M FMI51 can be installed from the top or from the bottom.
Make sure that:
• the probe is not installed in the area of the filling curtain
• the probe is not in contact with the container wall
• the distance from the container floor is ≥10 mm (0.39 in)
• multiple probes are mounted next to each other at the minimum distance between
the probes of 500 mm (19.7 in)
• the probe is at a sufficient distance from the agitator if using the probe in agitator
tanks
• the rod probes with a ground tube are used in the event of severe lateral load
A0040392
Unit of measurement mm (in)