12.6.3 Pressure and temperature derating
For process connections ½", ¾", 1", flanges <DN50, <ANSI 2", <JIS 10K
(10 mm (0.39 in) rod) and process connections ¾", 1", flanges <DN50, <ANSI 2", <JIS
10K (16 mm (0.63 in) rod)
Rod insulation: PTFE, PFA
bar
P
p
0
–40
40
20
–80
+200/
–1
+200/
+2
+50/
+25
T
P
°C
40 80 120 160
A0040625
P
p
Process pressure
T
p
Process temperature
For process connections 1½", flanges <DN50, <ANSI 2", <JIS 10K (16 mm (0.63 in)
rod)
Rod insulation: PTFE, PFA
bar
P
p
0
–40
60
40
80
20
–80
+200/
–1
+200/
+20
+80/
+60
+40/
+100
T
P
°C
100
63
40 80 120 160
A0040626
P
p
Process pressure
T
p
Process temperature
63 Process pressure for probes with an inactive length
With a fully insulated inactive length (22 mm (0.87 in) rod
Rod insulation: PTFE, PFA