Commissioning Liquicap M FMI51 HART
74 Endress+Hauser
Editing the probe and application-specific data
Editing the probe and application-specific data.
1.
Click the "Probe type" button.
2. Select the probe type.
3. Enter the length probe L1 in accordance with the nameplate.
4. Enter the inactive length L3 in accordance with the nameplate.
5. Enter the "Value empty E".
6. Enter the "Value full F".
7. Enter the "Wall distance".
8. Enter the conductivity value of the medium in the "Medium top".
9. Enter the dielectric constant value of the medium in the "Medium top".
10. Enter the conductivity value of the medium in the "Medium bottom".
11. Enter the dielectric constant value of the medium in the "Medium bottom".
12. Click the "Calibration data interface measurement" button to obtain the capacitance
values for the calibration.
The capacitance values for empty calibration and full calibration are calculated
and displayed as a result.
Use the "DC handbook" button to transfer the DC values and the conductivity of the
corresponding media to the calculation program, if the medium properties are not
known.
7.9.4 Wet calibration for interface measurement
This chapter describes the wet calibration procedure for "Empty calibration" and "Full
calibration".
Empty calibration
1. Fill the container with the top medium.