Liquicap M FMI51 HART Mounting
Endress+Hauser 21
H5 parameter
• probes ⌀10 mm (0.39 in) rod
66 mm (2.6 in)
• probes ⌀16 mm (0.63 in) rod, without fully insulated inactive length
• 66 mm (2.6 in)
• threads: G¾", G1", NPT¾", NPT1", Clamp 1", Clamp 1½", Universal ⌀44 mm (1.73 in),
flange < DN50, ANSI 2", 10K50
• probes ⌀16 mm (0.63 in) rod, without fully insulated inactive length
• 89 mm (3.5 in)
• threads: G1½", NPT1½", Clamp 2", DIN 11851, flange ≥ DN50, ANSI 2", 10K50
• probes ⌀22 mm (0.87 in) rod, with fully insulated inactive length
89 mm (3.5 in)
4.6.2 Wall bracket
The wall bracket is part of the scope of delivery.
The wall bracket must be screwed first to the separate housing before you can use it
as a drilling template.
The distance between the holes is reduced by screwing it to the separate housing.
28 (1.1)
86 (3.39)
70 (2.76)
6.2 (0.24)
3 (0.12)
A0033881
Unit of measurement mm (in)
4.6.3 Wall mounting
1 → 2 → 3
A0042318 A0042319 A0042320