Mounting Liquicap M FMI51 HART
22 Endress+Hauser
4.6.4 Pipe mounting
The maximum pipe diameter is 50.8 mm (2 in).
1 → 2
A0042318 A0042321
4.6.5 Shortening the connecting cable
NOTICE
Risk of damage to connections and cable.
‣
Make sure that neither the connecting cable nor the probe is turning with the pressing
screw!
Recalibration must be performed before commissioning.
The maximum connection length between the probe and the separate housing is
6 m (20 ft).
When ordering a device with separate housing, the desired length must be specified.
If the cable connection has to be shortened or led through a wall, it must be separated
from the process connection.
Disconnecting the connection cable
1 → 2 → 3 →
A0042111 A0042112 A0042113
4 → 5 → 6 →
A0042114 A0042115 A0042116