Liquicap M FMI51 HART Electrical connection
Endress+Hauser 27
Cable entries
• nickel-plated brass
⌀d = 7 to 10.5 mm (0.28 to 0.41 in)
• synthetic material
⌀d = 5 to 10 mm (0.2 to 0.38 in)
• stainless steel
⌀d = 7 to 12 mm (0.28 to 0.47 in)
5.1.4 Connector
For the version with a connector M12, the housing does not have to be opened for
connecting the signal line.
PIN assignment for M12 connector
A0011175
1 Positive potential
2 Not used
3 Negative potential
4 Ground
5.1.5 Supply voltage
All of the following voltages are terminal voltages directly at the device:
• 12.0 to 36.0 V
DC
in the non-hazardous area
• 12.0 to 30.0 V
DC
in the Ex ia hazardous area
• 14.4 to 30.0 V
DC
in the Ex d hazardous area
5.2 Wiring and connecting
5.2.1 Connection compartment
Depending on explosion protection, the connection compartment is available in the
following variants:
Standard protection
• polyester housing F16
• stainless steel housing F15
• aluminum housing F17
• aluminum housing F13
• stainless steel housing F27
• aluminum housing T13, with the separate connection compartment
Ex ia protection
• polyester housing F16
• stainless steel housing F15
• aluminum housing F17
• aluminum housing F13
• stainless steel housing F27
• aluminum housing T13, with the separate connection compartment