Liquicap M FMI51 HART Mounting
Endress+Hauser 23
7 → 8 → 9
A0042117 A0042118 A0042119
gn/ye
ye
rd
bk
6
1
2
3
4
5
A0040734
 9 Cable connections
1 External screening (not required)
2 Strand black (bk) (not required)
3 Coaxial cable with central core a screening
4 Solder the red (rd) strand with the central core of the coaxial cable (probe)
5 Solder the strand with the screening of the yellow (ye) coaxial cable (ground)
6 Strand yellow and green (gn/ye) with a ring terminal
We recommend reusing all strands with ring terminals in case of shortening the
connecting cable.
To avoid the risk of short-circuiting when the strands are not to be reused, the
connections of the new ring terminals fitted must be isolated with a heat shrinking
sleeve.
Use heat-shrink tubes to insulate all soldered joints.
4.7 Installation instructions
NOTICE
Do not damage the probe insulation during installation!
Check the rod insulation.
NOTICE
Do not screw the probe using the probe housing!
Use an open-end wrench to screw the probe.