Liquicap M FMI51 HART Mounting
Endress+Hauser 23
7 → 8 → 9
A0042117 A0042118 A0042119
A0040734
9 Cable connections
1 External screening (not required)
2 Strand black (bk) (not required)
3 Coaxial cable with central core a screening
4 Solder the red (rd) strand with the central core of the coaxial cable (probe)
5 Solder the strand with the screening of the yellow (ye) coaxial cable (ground)
6 Strand yellow and green (gn/ye) with a ring terminal
We recommend reusing all strands with ring terminals in case of shortening the
connecting cable.
To avoid the risk of short-circuiting when the strands are not to be reused, the
connections of the new ring terminals fitted must be isolated with a heat shrinking
sleeve.
Use heat-shrink tubes to insulate all soldered joints.
4.7 Installation instructions
NOTICE
Do not damage the probe insulation during installation!
‣
Check the rod insulation.
NOTICE
Do not screw the probe using the probe housing!
‣
Use an open-end wrench to screw the probe.