Liquicap M FMI51 HART Commissioning
Endress+Hauser 49
L3 for G1½ = 25 mm (0.98 in)
L3 for G <1½ = 19 mm (0.75 in)
Plug:
10 mm (0.39 in) rod = 10 mm (0.39 in)
16 mm (0.63 in) rod = 15 mm (0.59 in)
22 mm (0.87 in) rod = 15 mm (0.59 in)
Subfunction: "Cap. empty"
The calculated capacitance value is displayed here. This field cannot be edited.
Subfunction: "Confirm cal."
Empty calibration is confirmed with this subfunction.
7.3.7 Operating mode: "Full calibration" and function - "Dry" for
conductive and nonconductive media
The "Full" value can be entered directly in units of length.
Subfunction: "Value full", medium property - conductive, nonconductive
4 mA
20 mA
0 %
100 %
F
L1
L3
A0040523
L1 Active probe rod.
L3 Thread length.
F The distance from the zero-point to the desired 100% point.
Value full:
F ≤ Empty value
Subfunction: "Cap. full"
The calculated capacitance value is displayed here. This field cannot be edited.
Subfunction: "Confirm cal."
Full calibration is confirmed with this subfunction.