4.6 Probe with separate housing
A0040466
7 Connection of the probe and separate housing
A Explosive zone 1
B Explosive zone 0
L1 Rod length: max. 4 m (13 ft)
L4 Cable length
The maximum cable length L4 and rod length L1 cannot exceed 10 m (33 ft).
The maximum cable length between the probe and separate housing is 6 m (20 ft).
The required cable length must be indicated in the ordering process of a Liquicap M
with separate housing.
If the cable connection has to be shortened or led through a wall, then it must be
separated from the process connection.
4.6.1 Extension heights: separate housing
The cable has:
• a minimum bending radius of r ≥ 100 mm (3.94 in)
• diameter ⌀ 10.5 mm (0.14 in)