Mounting Liquicap M FMI51 HART
20 Endress+Hauser
61 (2.4)
41 (1.61)
B
H1
75 (2.95)
41 (1.61)
B
H3
H1
H5
D
r 100
r 100
A0040471
 8 Housing side: wall mounting, pipe mounting, and sensor side. Unit of measurement mm (in)
The parameters value
1)
:
B parameter
polyester housing (F16)
76 mm (2.99 in)
stainless steel housing (F15)
64 mm (2.52 in)
aluminum housing (F17)
65 mm (2.56 in)
H1 parameter
polyester housing (F16)
172 mm (6.77 in)
stainless steel housing (F15)
166 mm (6.54 in)
aluminum housing (F17)
177 mm (6.97 in)
D parameter
probes ⌀10 mm (0.39 in) rod
38 mm (1.5 in)
probes ⌀16 mm (0.63 in) rod, without fully insulated inactive length
38 mm (1.5 in)
threads: G¾", G1", NPT¾", NPT1", Clamp 1", Clamp 1½", Universal ⌀44 mm (1.73 in),
flange < DN50, ANSI 2", 10K50
probes ⌀16 mm (0.63 in) rod, without fully insulated inactive length
50 mm (1.97 in)
threads: G1½", NPT1½", Clamp 2", DIN 11851, flange ≥ DN50, ANSI 2", 10K50
Probes ⌀22 mm (0.87 in) rod, with fully insulated inactive length
38 mm (1.5 in)
1) See parameters on the drawings.