Liquicap M FMI51 HART Commissioning
Endress+Hauser 75
2. Perform "Empty calibration 0%" regarding procedure →  44.
If it is not possible to fill the tank, perform the "Empty calibration" with the probe
exposed in the air, however expect a calibration inaccuracy approximately 2.5 % per
meter. The water and oil are the reference media.
Full calibration
1. Fill the container with the bottom medium.
2. Perform "Full calibration 100%" regarding procedure →  44.
The empty and full calibration is completed and all data are saved in the electronic insert
and DAT sensor.