Liquicap M FMI51 HART Mounting
Endress+Hauser 17
If the probe is installed in a plastic tank, then a probe with the ground tube must be used.
The probe housing must be grounded.
A fully insulated rod probe may be neither shortened nor extended.
Damaged insulation of the probe rod causes improper measurements.
The following application examples show the vertical installation for continuous level
measurement.
A0040425
 2 A probe with the conductive tanks
A0040426
 3 A probe with the ground tube for the nonconductive tanks
A0040427
 4 A probe with the inactive length for the insulated tanks