Liquiline System CA80SI Diagnostics and troubleshooting
Endress+Hauser 115
No. Message Factory settings Tests or remedial action
S
1)
D
2)
F
3)
337 Pump hose warning M On Off End of pump hose service life will be reached
shortly
Displayed under Menu/Diagnostics/Operating
time information/Pump tube life
1. Schedule replacement
2. After replacement, reset the operating time
under Menu/Diagnostics/Operating time
information
360 Cooling/Heating C On Off Temperature range in the housing exceeded
1. Check installation conditions and the
ambient temperature
2. Replace the FMAB1 module
3. Contact the Service Department
361 Cooling/Heating F On On Cooling/heating module defective
Defined temperature range not reached. This
could affect the functionality of the reagents.
1. Check whether the insulation cap is seated
correctly over the reagents
2. Replace cooling/heating module
3. Contact the Service Department
362 Photometer temp. F On Off Temperature of the photometer too high
Contact the Service Department
363 Photometer temp. F On Off Temperature of the photometer too low
Contact the Service Department
365 Photometer comm. F On On Photometer not communicating
Possible reasons:
Incorrect photometer connection
1. Check photometer connection
2. Contact the Service Department
367 Module connect. F On On No communication to sample preparation
Check connecting cable to sample preparation
system
370 Internal voltage F On On Internal voltage outside the valid range
Check supply voltage
373 Electronictemp. high M On Off High electronics temperature
Check ambient temperature and energy
consumption
374 Sensor check F On Off No measurement signal from sensor
1. Check sensor connection
2. Check sensor, replace if necessary
380 Firmware failure F On On Internal software error
1. Update the software
2. Replace backplane
3. Contact the Service Department and quote
the number displayed
385 Heating temp. low S On Off Sample pre-heating target temperature not
reached
1. Check fuse.
2. Check cable and connector.