Diagnostics and troubleshooting Liquiline System CA80SI
116 Endress+Hauser
No. Message Factory settings Tests or remedial action
S
1)
D
2)
F
3)
386 Heating temp. high S On Off Temperature of sample pre-heating too high
1. Monitor the temperature of the sample.
2. Disconnect the power supply if the sample
temperature does not drop.
401 Factory reset F On On Factory reset is performed
405 Service IP active C Off Off Service switch is switched on
The device can be addressed at 192.168.1.212.
Switch off the service switch to change to the
saved IP settings
406 Param. active C Off Off
Wait for configuration to be finished
407 Diag. setup active C Off Off
Wait for maintenance to be finished
412 Writing backup F On Off
Wait for the write process to be finished
413 Reading backup F On Off
Wait
460 Output below limit S On Off Reasons
Sensor in air
Air pockets in assembly
Sensor fouled
Incorrect flow to sensor
1. Check sensor installation
2. Clean sensor
3. Adjust assignment of current outputs
461 Output above limit S On Off
502 No text catalog F On On
Contact the Service Department
503 Language change M On Off Language change failed
Contact the Service Department
529 Diag. setup active C Off Off
Wait for maintenance to be finished.
530 Logbook at 80% M On Off 1. Save the logbook to the SD card and then
delete the logbook in the device
2. Set memory to ring memory
3. Deactivate logbook
531 Logbook full M On Off
532 License error M On Off
Contact the Service Department
540 Parameter save fail M On Off Storage of configuration has failed
Repeat
541 Parameter load ok M On Off Configuration successfully loaded
542 Parameter load fail M On Off Loading of configuration has failed
Repeat
543 Parameter load abort M On Off Configuration loading aborted
544 Parameter reset ok M On Off Factory default successful
545 Parameter reset fail M On Off Setting of device configuration to factory setting
has failed
565 Configuration M On Off Invalid configuration of sample preparation
system
1. In Setup/Sample preparation check the
number of channels used and their
operating mode and installation type.
2. Check the permitted combinations of
sample preparation and analyzers, see the
Operating Instructions for the sample
preparation system.