4-/6-channel device: Drain sample channels
1. Disconnect sample inlet hose SPx from the sample channel switching: Disconnect
sample inlet hose SPx from the T-section downstream from the sample channel
switching.
2. Inflate the maintenance dispenser with air.
3. Slide the C-Flex hose of the maintenance dispenser onto the end of sample inlet hose
SPx.
4. Drive air through the sample channel.
Any liquid remaining in the sample channel is driven into the cuvette.
5. Repeat steps 2 to 4 until the sample channel is empty.
Emptying the cuvette
1. Remove the cover in front of the cuvette.
2. Insert the maintenance dispenser together with the hose piece provided through the
hole in the capillary holder of the cuvette.
3. Empty the cuvette completely.