Liquiline System CA80SI Operation
Endress+Hauser 55
3. Set the recording rate (Scan time).
You can set the recording rate individually for every data logbook.
Further information on the logbooks: .
Data logbooks only apply for sensors (optional). There are special data logbooks for
the analyzer. These logbooks are automatically enabled and assigned to the
measuring channel.
SP1: Data logbook SP1 is assigned to measuring channel 1 of the analyzer.
Menu/Setup/General settings/Logbooks
Function Options Info
Logbook ident Customized text, 16
characters
Part of the file name when exporting a logbook
Event logbook Selection
Off
Ring buffer
Fill up buffer
Factory setting
Ring buffer
All diagnostic messages are recorded
Ring buffer
If the memory is full, the most recent entry
automatically overwrites the oldest entry.
Fill up buffer
If the memory is full, there is an overflow,i. e.
you cannot store any new values. The controller
displays a corresponding diagnostic message.
The memory then has to be cleared manually.
Analyzer event logbook Selection
Ring buffer
Fill up buffer
Factory setting
Ring buffer
All diagnostic messages are recorded
Ring buffer
If the memory is full, the most recent entry
automatically overwrites the oldest entry.
Fill up buffer
If the memory is 80 % full, the device displays a
diagnostic message.
If the memory is full, there is an overflow, i. e. no
new values can be saved. The controller displays
a corresponding diagnostic message. The
memory then has to be cleared manually.
Overflow warnings
Analyzer calibration logbook
= Fill up buffer
Calibration logbook Selection
Off
On
Factory setting
Off
Decide whether you want to receive a
diagnostic message if the fill buffer of the
relevant logbook overflows.
Diagnostic logbook
Configuration logbook
Analyzer data logbooks For analyzer measured data
The entry is made automatically on completion
of the measurement. A setting does not need to
be made. The logbook is activated automatically.
SP1 is assigned to data logbook SP1.