Operation Liquiline System CA80SI
66 Endress+Hauser
11.2.1 Advanced setup
Menu/Setup/Analyzer/Extended setup
Function Options Info
Measuring value
Main value Selection
SiO2
SiO2-Si
H2SiO3
Factory setting
SiO2
Unit Selection
mg/l
µg/l
ppb
Factory setting
µg/l
Main value format Selection
#
#.#
#.##
#.###
Auto
Factory setting
Auto
Signal for process access
SP1 Selection
Always accessible
Binary input x:y
Factory setting
Always accessible
Always accessible: The input signal level at the
binary inputs does not affect actions that require
sample (measurement, calibration, cleaning).
Binary input x:y: When the signal is active at
the selected input, the analyzer starts actions
where sample is absolutely required. Otherwise
the analyzer defers or skips the action.
With multi-channel devices, SP2 to SP6is also
displayed, depending on the number of
channels.
Diagnostics settings
Limits dispensers
Control Selection
Off
On
Factory setting
On
Warning limit
Remaining
operating time
Read only
Dispenser 1
Dispenser 2,
Dispenser 3,
Selection
1 to 90 (d)
Factory setting
28 d
Diag. code 733 Read only
Alarm limit
Remaining
operating time
Selection
1 to 60 (d)
Factory setting
7 d