Operation Liquiline System CA80SI
66 Endress+Hauser
11.2.1 Advanced setup
Menu/Setup/Analyzer/Extended setup
Function Options Info
Measuring value
Main value Selection
• SiO2
• SiO2-Si
• H2SiO3
Factory setting
SiO2
Unit Selection
• mg/l
• µg/l
• ppb
Factory setting
µg/l
Main value format Selection
• #
• #.#
• #.##
• #.###
• Auto
Factory setting
Auto
Signal for process access
SP1 Selection
• Always accessible
• Binary input x:y
Factory setting
Always accessible
Always accessible: The input signal level at the
binary inputs does not affect actions that require
sample (measurement, calibration, cleaning).
Binary input x:y: When the signal is active at
the selected input, the analyzer starts actions
where sample is absolutely required. Otherwise
the analyzer defers or skips the action.
With multi-channel devices, SP2 to SP6is also
displayed, depending on the number of
channels.
Diagnostics settings
Limits dispensers
Control Selection
• Off
• On
Factory setting
On
Warning limit
Remaining
operating time
Read only
Dispenser 1
Dispenser 2,
Dispenser 3,
Selection
1 to 90 (d)
Factory setting
28 d
Diag. code 733 Read only
Alarm limit
Remaining
operating time
Selection
1 to 60 (d)
Factory setting
7 d