Liquiline System CA80SI Operation
Endress+Hauser 67
Menu/Setup/Analyzer/Extended setup
Function Options Info
Diag. code 732 Read only
Bottles
Control Selection
Off
On
Factory setting
Off
Standard S1 Selection
100 to 1000 ml
Factory setting
Warning limits
Standard S1 Selection
1 to 20 %
Factory setting
5 %
Reagent RB Selection
1 to 40 %
Factory setting
10 %
Reagent RK Selection
1 to 40 %
Factory setting
10 %
Reagent RN Selection
1 to 40 %
Factory setting
10 %
Diag. code 726 Read only
Alarm limits
Standard S1 Selection
1 to 20 %
Factory setting
2 %
Reagent RB Selection
1 to 40 %
Factory setting
5 %
Reagent RK Selection
1 to 40 %
Factory setting
5 %
Reagent RN Selection
1 to 40 %
Factory setting
5 %