Liquiline System CA80SI Operation
Endress+Hauser 69
Menu/Setup/Analyzer/Measurement
Function Options Info
Sequence of measurements Selection
• SP1
• SP2
• SP3
• SP4
• SP5
• SP6
• Pause
Defines the sequence of channels in automatic
mode.
Is displayed only in the case of a two-, four- or
six-channel device.
• SP1: Indicate the number of consecutive
measurements for channel SP1
• SP2: Indicate the number of consecutive
measurements for channel SP2
• SP3: Indicate the number of consecutive
measurements for channel SP3
• SP4: Indicate the number of consecutive
measurements for channel SP4
• SP5: Indicate the number of consecutive
measurements for channel SP5
• SP6: Indicate the number of consecutive
measurements for channel SP6
• Pause: If the Pause channel is selected, the
analyzer does not run the measurement.
Using the soft keys INSERT, DEL and SAVE you
can add rows to the table, delete and save them.
Signal delay 0 to 600
Factory setting
0
Here you can delay the process of outputting the
"Measurement active" signal when a
measurement is running by the specified time.
Measuring is paused for the duration of the
defined signal delay time.
11.2.3 Calibration
Menu/Setup/Analyzer/Calibration
Function Options Info
Start condition Selection
• Immediate
• Date/time
• Disabled
Factory setting
Immediate
The calibration can start either immediately or
on the set date/at the set time.
If the start condition selected is Immediate
Date 01.01.1970 to 07.02.2106
Factory setting
DD.MM.YYYY
Time 00:00:00 to 23:59:59
Factory setting
HH:MM:SS (24 h)
Calibration interval Factory setting
02-00
For configuring the time interval of the
calibration/adjustment.
Next calibration
Mode = Automatic
Read only
Zero point Read only
Calibration factor Read only Relationship of the measured concentration to
the pre-specified concentration of the calibration
standard.